CCMA · EKG and Cardiovascular Testing · 5 min read

Recording, Transmitting and Maintaining EKG Equipment

In this lesson you will
  • Identify standard EKG machine settings and recognize common recording artifacts along with their corrective troubleshooting actions.
  • Describe the medical assistant's scope of practice regarding tracing inspection, patient safety, documentation, and secure transmission.
  • Outline the proper maintenance, cleaning, storage, and handling protocols for EKG machines, cables, electrodes, and recording paper.

Standard Machine Settings and Patient Preparation

Before obtaining an electrocardiogram (EKG), the medical assistant must verify that the machine is configured to standard settings. The standard paper speed for a resting 12-lead EKG is 25 mm per second, and the standard amplitude or gain setting is 10 mm per millivolt. When the machine prints, it generates a standard calibration mark that measures 10 mm in height and 5 mm in width. These baseline parameters ensure that providers can accurately evaluate the timing and size of the heart's electrical events.

Patient positioning directly influences the quality of the tracing and patient comfort. Position the patient supine for a resting EKG whenever it is tolerated. However, if the patient exhibits signs of respiratory distress, allow a more upright position to ease breathing and promptly alert the provider. Once positioned, instruct the patient to lie still, breathe normally, and avoid talking during the recording, as any movement can disrupt the electrical signal.

Occasionally, unique patient circumstances require a change in the machine settings. A provider may request increasing the paper speed to 50 mm per second to spread out waveforms during extreme tachycardia. Conversely, if QRS complexes are exceptionally tall and overlap each other, the gain setting may be adjusted to half-standard at 5 mm per millivolt. Any intentional alteration from standard speed or gain settings must be noted directly on the EKG tracing so the provider is aware of the modification during evaluation.

Exam tip. Look out for exam questions that describe overlapping, tall waveforms or extreme rapid heart rates. Remember that half-standard gain is 5 mm per millivolt, and double paper speed is 50 mm per second. Always note any setting changes directly on the printout.

Common mistakes

  • Disconnecting electrodes from the patient before checking the tracing for quality, which requires reapplying fresh electrodes if artifacts make the tracing unreadable.
  • Attempting to interpret an abnormal rhythm or ST-segment change instead of immediately alerting the provider.
  • Leaving an unstable or symptomatic patient alone in the exam room to locate the provider or retrieve equipment.
  • Tightly wrapping or bending lead wires around the machine, which damages the internal conductors over time.

Key points

  • Standard resting 12-lead EKG parameters are a paper speed of 25 mm/sec, a gain of 10 mm/mV, and a calibration mark measuring 10 mm high by 5 mm wide.
  • Patients should be supine for resting tracings; if respiratory distress occurs, permit an upright position and alert the provider promptly.
  • Common artifacts include somatic tremor (muscle movement), wandering baseline (loose/dry electrodes or oils), AC interference (electrical devices), and interrupted baseline (broken or detached lead wires).
  • Medical assistants do not diagnose or interpret tracings; always stay with symptomatic patients and immediately alert the provider to critical findings.
  • Document at least two patient identifiers, date, time, any symptoms, and any setting changes directly on the tracing.
  • Transmit EKG data via HIPAA-compliant methods, verifying fax numbers and using secure cover sheets when faxing.
  • Maintain equipment by loosely coiling cables, disinfecting per standard precautions without unauthorized autoclaving or immersion, properly storing thermal paper, and discarding single-use or expired electrodes.

Practice this section

Explained CCMA EKG and Cardiovascular Testing questions, no account needed. CCMA EKG and Cardiovascular Testing practice test

Study guide included with CCMA plans

The full lesson continues with 3 more sections with exam tips, audio in English and Spanish, 5 Check yourself questions and 20 flashcards.

  • 1 month$41.99 $20.99
  • 3 months$122.99 $61.49
  • Lifetime$251.99 $125.99

One payment, no subscription.

See CCMA plans

Already a member? Log in

Sample lesson: Electrodes, Leads and 12-Lead Placement

25 practice questions on this exam

Each CCMA practice test has 25 original questions with the answers explained. No account needed.

Open the CCMA practice tests