CCMA · Testing and Laboratory Procedures · 8 min read
Vision, Hearing, Allergy and Pulmonary Function Tests
- Explain the step-by-step procedures for conducting distance, near, and color visual acuity screenings using standardized charts and testing distances.
- Describe clinical workflows for audiometric testing, tuning fork assessments, and safe ear irrigation while identifying methods to prevent caloric complications.
- Outline the medical assistant's responsibilities during allergy skin tests, spirometry, and peak flow monitoring, focusing on patient safety protocols, preparation instructions, and documentation.
Visual Acuity and Color Perception Screening
In the outpatient setting, the medical assistant routinely conducts visual acuity screenings under provider supervision during rooming or routine examinations. Distance visual acuity is most commonly assessed using a standard Snellen chart. The test is conducted with the patient standing or sitting exactly 20 feet away from the chart. The medical assistant instructs the patient to keep both eyes open throughout the procedure while using an opaque occluder to cover one eye at a time. Testing is typically performed on the right eye (OD) first, followed by the left eye (OS), and finally both eyes together (OU). If the patient wears corrective lenses for distance vision, the medical assistant tests and records their visual acuity with the lenses on first, unless the provider specifically instructs otherwise.
The results of a Snellen test are expressed as a fraction. The numerator represents the distance the patient stands from the chart, which is standard at 20 feet. The denominator represents the distance at which an individual with normal vision can read that specific line of characters. A patient successfully passes a line on the Snellen chart if they correctly identify more than half of the characters displayed on that line. For preliterate pediatric patients or non-English-speaking patients who may struggle with standard Latin letters, the medical assistant uses an alternative screening tool, such as the Tumbling E chart or the Landolt C chart.
Screening for near visual acuity assesses the patient's ability to read material close at hand, helping to detect conditions such as hyperopia and presbyopia. The medical assistant conducts this screening using a handheld card, such as a Jaeger card or a Rosenbaum pocket vision screener. The patient holds the test card 14 to 16 inches away from their eyes, similar to the natural distance one might hold a book or a smartphone. Each eye is tested individually and then both eyes together, documenting the findings accurately for the provider's review.
Color vision screening identifies inherited or acquired color vision deficiencies, particularly red-green color vision deficiencies. The medical assistant administers the Ishihara test by presenting test plates approximately 30 inches (75 cm) from the patient's eyes. The plates must be held perpendicular to the patient's line of sight and viewed under natural daylight or standardized illumination to avoid distorted color perception. The medical assistant allows the patient approximately 3 seconds to identify the number or trace the pattern present on each plate before documenting the responses.
| Screening Tool | Clinical Purpose | Standard Testing Distance | Special Patient Considerations |
|---|---|---|---|
| Snellen Chart | Distance visual acuity | 20 feet | Test right eye (OD), left eye (OS), then both (OU); keep corrective lenses on first unless instructed otherwise. |
| Tumbling E / Landolt C | Distance visual acuity | 20 feet | Used for preliterate pediatric patients or non-English-speaking patients. |
| Jaeger Card / Rosenbaum Screener | Near visual acuity (screens for hyperopia and presbyopia) | 14 to 16 inches | Held at customary reading distance; patient covers one eye at a time while keeping both eyes open. |
| Ishihara Plates | Screens for red-green color vision deficiencies | 30 inches (75 cm) | Presented perpendicular to line of sight under natural daylight or standardized illumination; 3 seconds per plate. |
Common mistakes
- Allowing a patient to close or squint an eye during a Snellen test instead of keeping both eyes open behind an opaque occluder, which artificially distorts distance acuity results.
- Using cold or unheated water for ear irrigation, which triggers the caloric effect and induces acute dizziness, nausea, or nystagmus in the patient.
- Averaging the three peak flow attempts rather than identifying and recording the single highest reading of the three acceptable maneuvers.
- Instructing patients to discontinue their prescribed antihistamines or bronchodilators without specific instructions from the supervising provider.
Key points
- Distance vision is screened at 20 feet using the Snellen chart (or Tumbling E/Landolt C for non-English-speaking or preliterate pediatric patients), while near vision is screened at 14 to 16 inches using a Jaeger card or Rosenbaum screener to detect hyperopia and presbyopia.
- The Ishihara test screens for red-green color deficiencies at 30 inches under daylight or standardized illumination, allowing approximately 3 seconds per plate.
- Audiometry tests sound frequencies (pitch) and decibels (intensity) using red headphones for the right ear and blue for the left; ear irrigation solution must be warmed to body temperature to prevent caloric-effect dizziness, nausea, and nystagmus.
- Allergy testing uses percutaneous scratch or delegated intradermal methods (10- to 15-degree angle), measured at 15 to 20 minutes using histamine (positive) and saline/glycerin (negative) controls, with emergency supplies (epinephrine, oxygen, antihistamines) ready for anaphylaxis.
- Spirometry measures FVC and FEV1 while the patient sits upright with feet flat and wearing a nose clip; at least three acceptable forced expiratory maneuvers meeting repeatability criteria are required.
- Peak expiratory flow rate evaluates airway constriction using a meter reset to zero; the patient performs three maneuvers, and the medical assistant records the highest of the three acceptable readings.
- Peak flow zones guide asthma management: Green (80%-100%), Yellow (50%-79%), and Red (<50%, requiring immediate provider notification and emergency protocol activation); all reusable devices are cleaned between uses, and results are documented without independent interpretation.
Practice this section
Explained CCMA Testing and Laboratory Procedures questions, no account needed. CCMA Testing and Laboratory Procedures practice test
The full lesson continues with 4 more sections with exam tips, audio in English and Spanish, 5 Check yourself questions and 20 flashcards.
- 1 month
$41.99$20.99 - 3 months
$122.99$61.49 - Lifetime
$251.99$125.99
One payment, no subscription.
See CCMA plansAlready a member? Log in
Sample lesson: Urine Collection, Urinalysis, Pregnancy and Drug Screens25 practice questions on this exam
Each CCMA practice test has 25 original questions with the answers explained. No account needed.
Open the CCMA practice testsOriginal lesson written for CCMA practice · updated October 31, 2026. Exam names are trademarks of their owners; tests.global is not affiliated with them.