86,762 verified 4-option questions Double-checked by independent solution

CRCST · Sterilization Process · 7 min read

Air Removal Tests, Biological Indicators and PCDs

In this lesson you will
  • Explain the purpose, procedure, and interpretation of dynamic air removal testing in pre-vacuum steam sterilizers.
  • Differentiate between biological indicator spore types, incubation technologies, and the roles of test and control indicators.
  • Describe the function of Process Challenge Devices (PCDs), load release requirements for implants, routine monitoring frequencies, and sterilizer qualification protocols.

Dynamic Air Removal Testing: Bowie-Dick and DART

In pre-vacuum steam sterilizers, air must be actively removed from the chamber so that steam can contact every surface of the items being processed. A Bowie-Dick test evaluates the efficacy of dynamic air removal in pre-vacuum steam sterilizers. This test detects air leaks, inadequate air removal, and the presence of non-condensable gases that can prevent steam from sterilizing instruments. The Daily Air Removal Test, or DART, is a commercially manufactured equivalent of the Bowie-Dick test. Technicians should note that Bowie-Dick testing is not applicable to gravity-displacement steam sterilizers, which rely on passive air displacement rather than mechanical dynamic air removal.

Air removal testing must be performed daily in dynamic air-removal steam sterilizers. A Bowie-Dick test should be run each day the sterilizer is used, typically before the first processed load of the day. When required by the sterilizer manufacturer's instructions for use, such as after the sterilizer has been shut down, a warm-up cycle must be run before the Bowie-Dick test. Before conducting the test, sterile processing technicians must verify the expiration dates of the Bowie-Dick test packs to ensure the chemical indicators are valid.

The Bowie-Dick test pack must be run in an otherwise completely empty sterilizer chamber. In a pre-vacuum steam sterilizer, the technician places the test pack horizontally over the drain on the bottom rack. After the cycle finishes, the test sheet inside the pack is evaluated. A uniform color change throughout the Bowie-Dick test sheet indicates a passing air removal test. If residual air is present in the chamber, it may produce a lighter or unchanged area, often near the center of the Bowie-Dick test sheet, indicating a failed test that must be addressed.

Exam tip. Always remember that a Bowie-Dick test must be run in a completely empty chamber, placed horizontally over the bottom rack drain, and is only used for dynamic air-removal steam sterilizers, never gravity-displacement units.

Common mistakes

  • Assuming an uncleaned instrument is sterile simply because the biological indicator in the load is negative; residual soil can shield microorganisms from sterilants.
  • Interpreting a negative control biological indicator as a pass; an unexposed control must always show positive growth to prove spore viability and incubator function.
  • Running a Bowie-Dick test with instrument sets inside the chamber; dynamic air removal tests must always be conducted in an otherwise completely empty chamber.
  • Releasing an implant load before biological indicator results are available without obtaining and documenting an authorized emergency release per facility policy.

Key points

  • Sterile processing technicians clean, decontaminate, inspect, assemble, package, sterilize, and store instruments following manufacturer instructions for use and facility policy.
  • Dynamic air removal tests (Bowie-Dick or DART) must be run daily in an empty chamber over the bottom rack drain to detect air leaks, inadequate air removal, and non-condensable gases in pre-vacuum steam sterilizers.
  • Biological indicators provide the only direct measure of cycle lethality, using Geobacillus stearothermophilus for steam, vaporized hydrogen peroxide, and ozone, and Bacillus atrophaeus for ethylene oxide and dry heat.
  • A processed test indicator should be negative for bacterial growth, while an unexposed control indicator from the same lot must be positive to validate spore viability and incubator conditions.
  • Monitor every implant load with an appropriate biological indicator PCD and hold it until a negative result is available, except for a documented emergency release; include a Class 5 integrating indicator for steam implant loads and follow method-specific requirements for other methods.
  • Qualification testing following installation, relocation, major repairs, or utility interruptions requires three consecutive Bowie-Dick tests and three consecutive BI PCD test cycles for each cycle type before returning dynamic air-removal steam sterilizers to service.

Practice this section

Explained CRCST Sterilization Process questions, no account needed. CRCST Sterilization Process practice test

Study guide included with CRCST plans

The full lesson continues with 4 more sections with exam tips, audio in English and Spanish, 5 Check yourself questions and 20 flashcards.

  • 1 month$44.99 $22.49
  • 3 months$131.99 $65.99
  • Lifetime$269.99 $134.99

One payment, no subscription.

See CRCST plans

Already a member? Log in

Sample lesson: Sterilizer Safety, Supplies and Daily Readiness

25 practice questions on this exam

Each CRCST practice test has 25 original questions with the answers explained. No account needed.

Open the CRCST practice tests
All CRCST lessons

Original lesson written for CRCST practice · updated October 14, 2026. Exam names are trademarks of their owners; tests.global is not affiliated with them.